| TCPSC Flourescence PL Measurement System |
|
|
|
|
|
(UV-VIS) TCSPC Flourescence + PL Measurement System
1. MPL-375-20 Picosecond Diode Laser |
- Wavelength range: 375 nm(+/-10 nm)
- FWHM: <10 nm
- Peak light output: 20 mW( typ.)
- Output stability: <1.5%
- Pulse widt(ps): 100( max.) 80(typ.)
- Picosecond laser driver |
| 2. SCP300 PL Sample Chamber |
- 50mm dia., fused silica focusing lens set
- Laser rejection & Bandpass filter (3-10 nm FWHM)
- Adjustable input beam steering mirror
- Cuvette cell sample holder set |
| 3. MonoRa 324i Monochromator |
- Stepping motor scanning system with 32-bit microprocessor control
- Power supply and Monochromator Instruction manual included
- Monoscan basic control software included
- Resolution: 0.09nm @ 435.8 nm (1200gr/mm grating) 10 um slit
- Focal length: 320mm
- Optical path: Czerny-Turner type, Imaging Spectrograph by Toroidal Mirrors.
- Aperture: f/4.2,
- Grating Turret: dual-grating (68 mm x 68 mm) or triple turret
- Interface: RS232 & USB Standard
- Accuracy: +/- 0.2nm
- Repeatability: +/- 0.04nm
- Drive-step size: 0.0025 nm with 1200gr/mm
- Dispersion: 2.5nm/mm
- Focal-plane size: 26mm wide x 14mm high
- Fixed diverter mirror assembly for entrance side port
- Exit selection diverter mirror assembly for exit ports
- Micro adjustable slit assembly for entrance and exit ports
0 to 5mm(10micrometer increment/decrement unit)
- 32 bit control board
- Gratings:
A) 1-180-500, 31800 G/mm, 500 blaze
330-900nm Optimal Wavelength Range |
| 4. PMT Single Photon Detector |
- Combined with monochromator directly
- Time resolution: <250ps IRF
- Wavelength range: 200-820nm
- Supports pulse frequency up to 80MHz
- Power supply |
| 5. Time-Correlated Single Photon Counting Module |
- Electrical time resolution: 10 ps (at 20 ns)
- Histogrammer measurement range: 20 ns, 100 ns, 200 ns, 2 us, 20 us
- Time resolution ability: 10 ps (at 40 ns)
- PC interface: USB
- Signal input: PMT input Trig output
- Pluse Frequency: supports 50, 25, 12.5 or 6.25 MHz
- Input voltage : AC100V +/-10V
- Power consumption : 50 V1 |
| 6. Lifetime Analysis Application Software |
- Time scale range: 16.9ns~1.08us
- Measuring time limit: <10hrs
- Excitation period : 15 usec
- Flexible rescale of lifetime area
- Flexible coefficient contol
- Supports approximated curve and De-convolution drawing
- Supports MS Excel compatible format (.csv) |
| 7. SI-010-1 UV Enhanced, non-cooled type Si photodetector |
- MonoRa series compaible mount
- 2.5mm2 active area
- Power supply
- Optimal detection rage: 200nm~1100nm |
| 8. DAD-1602P 16Bit AD/DA data acquisition board |
|
- USB interface |
| 9. Monoworks Data Acquisition & Processing Application Software |
- Compatible with MonoRa series Monochromator(s)
- MonoRa series Monochromator(s) control
- General spectroscopy applications |
| 10. Data Acquisition & Processor |
- Windows XP (English) and
- 17" TFT-LCD monitor |
|
|
|
|
 |
|
|
|
|